Almaarefa University
The authoritative source for research information and outputs is Elsevier Scopus, with researcher profile verification from the Deanship of Scientific Research.
ArticleOpen access2026SDG 30 citation

Long-standing inferior vena cava filter with multidirectional strut penetration: A case report

Nawaf Salah Ayad Mohamed ✉،Osama Osman Ahmed Farah،Mohammed Ibrahim Elhamaky،Abdulaziz Mubarak Almasan،AbdulRahman BinSwilim،Talal Mislat Alotaibi

Radiology Case Reports · vol. 21 (8) · 3467-3470 · 2026

Abstract

Inferior vena cava (IVC) filter penetration is an increasingly recognized complication, particularly after prolonged dwell time or failed retrieval. Although often asymptomatic, multidirectional penetration into adjacent structures may pose a significant risk. We report a 61-year-old man presenting with non-traumatic radicular back pain whose CT imaging demonstrated a retrievable IVC filter placed nine years earlier with extensive strut penetration into the L3 vertebral body, perivertebral soft tissues, bowel loops, distal abdominal aorta, and right mesocolon, along with one fractured strut. Prior retrieval attempts had failed due to filter tilt and hook embedding, without evidence of hemorrhage or retroperitoneal leakage. The patient was referred for complex endovascular retrieval. This case underscores the importance of structured surveillance, timely retrieval, and specialized multidisciplinary management to prevent severe delayed complications.

Details

Journal
Radiology Case Reports
Publisher
Elsevier BV
Volume
21
Issue
8
Pages
3467-3470
ISSN
1930-0433
Year
2026
Language
en
License
cc-by
Open access
gold
Citations
0
Source
scopus

Topics

MedicineInferior vena cava filterInferior vena cavaRadiologyVena cavaFilter (signal processing)

Vascular anomalies and interventions (Medicine) · Venous Thromboembolism Diagnosis and Management (Medicine) · Central Venous Catheters and Hemodialysis (Health Professions)

Cite this

APA
Mohamed, N. S. A., Farah, O. O. A., Elhamaky, M. I., Almasan, A. M., BinSwilim, A., Alotaibi, T. M. (2026). Long-standing inferior vena cava filter with multidirectional strut penetration: A case report. Radiology Case Reports, 21(8), 3467-3470. https://doi.org/10.1016/j.radcr.2026.04.060
Harvard
Mohamed, N. S. A., Farah, O. O. A., Elhamaky, M. I., Almasan, A. M., BinSwilim, A., Alotaibi, T. M. 2026, 'Long-standing inferior vena cava filter with multidirectional strut penetration: A case report', Radiology Case Reports, vol. 21(8), 3467-3470. Available at: https://doi.org/10.1016/j.radcr.2026.04.060
Vancouver
Mohamed NSA, Farah OOA, Elhamaky MI, Almasan AM, BinSwilim A, Alotaibi TM. Long-standing inferior vena cava filter with multidirectional strut penetration: A case report. Radiology Case Reports. 2026;21(8), 3467-3470.
BibTeX
@article{mohamed2026173,
  title = {Long-standing inferior vena cava filter with multidirectional strut penetration: A case report},
  author = {Nawaf Salah Ayad Mohamed and Osama Osman Ahmed Farah and Mohammed Ibrahim Elhamaky and Abdulaziz Mubarak Almasan and AbdulRahman BinSwilim and Talal Mislat Alotaibi},
  year = {2026},
  journal = {Radiology Case Reports},
  volume = {21},
  number = {8},
  pages = {3467-3470},
  publisher = {Elsevier BV},
  doi = {10.1016/j.radcr.2026.04.060},
  url = {https://doi.org/10.1016/j.radcr.2026.04.060}
}
RIS
TY  - JOUR
AU  - Nawaf Salah Ayad Mohamed
AU  - Osama Osman Ahmed Farah
AU  - Mohammed Ibrahim Elhamaky
AU  - Abdulaziz Mubarak Almasan
AU  - AbdulRahman BinSwilim
AU  - Talal Mislat Alotaibi
TI  - Long-standing inferior vena cava filter with multidirectional strut penetration: A case report
PY  - 2026
JO  - Radiology Case Reports
VL  - 21
IS  - 8
SP  - 3467-3470
DO  - 10.1016/j.radcr.2026.04.060
UR  - https://doi.org/10.1016/j.radcr.2026.04.060
ER  - 

Related outputs